ONLINE GAS ANALYZER EXPERTS
WWW.CHROMATOTEC.COM
MED O R ® Exd ATEX
On-line analysis & monitoring of sulfur compounds in natural gas
and gaseous fuels for hazardous area
TSP MEDOR Exd ATEX
GAS ANALYSER GC 866
Chromatotec ® is specialised in VOC, Sulfur and permanent gases analysis at trace and ultra trace levels (ppm, ppb, ppt). Please visit our website for more details. Updated: February 2021
Model: Medor ® Exd ATEX
Main applications:
Impurities detection in Natural Gas / LPG / Propane / Butane
Propellant gas
Catalyzer protection
Targetted compounds:
In standard : THT / H 2S / DMS / Mercaptans: MM / EM / IPM / TBM /NPM
In option: 2 BM / IBM / NBM
Main markets:
Petrochemical
Gas transportation
Process
Standard:
ASTM D7493-08, ISO 19739:2004, DIN 51855/7
THT
TBM
DMS
H2S
Bu-SH
Me-SH
Et-SH
ATEX or IECEx:
Zone 1 and 2 group
IIB + H2
WWW.CHROMATOTEC.COM
Chromatotec ® is continuously improving its products, therefore these specifications are subject to change without notice
MED O R ® Exd ATEX
On-line analysis & monitoring of sulfur compounds in
natural gas and gaseous fuels for hazardous area
To contact us: sales@chromatotec.com
NORTH AMERICAHouston - USA EUROPEBordeaux - FRANCE
ASIA Beijing - CHINA
Options:
• External multiple stream selector (up to 16 streams controlled by
the analyzer)
• Calculation modules (Average / Statistics / Odor index...)
• Electric selection valve to reduce air consumption
• 24 V DC power supply, can work on battery or solar panel
• Inbuilt nitrogen generator for hazardous area from instrument air
or from external compressor
• Liquid Injection valve for LPG or up to condensate analysis
• Sampling vavle for high concentration up to % level
• Internal temperature control system using electric heater / VOR -
TEX / electric peltier cooler
To order:MEDOR ® Exd Atex zone1 MEDOR ® Exd Atex zone2
Technical specifications:
• Speciation and/or total sulfur
Detection limits:
• MEDOR® Exd ppm: • H2S: 0,1 ppm (0,14 mg/m3) • MEDOR ® Exd ppb:
• H2S: 5 ppb (7,0 µg/m3) or DMS: 2 ppb (5,1 µg/m3)
Range adjustable, depending on application:
• 0/10 or 0/100 or 0/1000 ppm or ppb• Calculation: total sulfur, total mercaptans...
Relative Standard Deviation:
• RSD < 3% on concentration over 48H.
• RSD < 0.5% on retention time over 48H.
Cycle Time:
• H2S 120 s• H2S/TOS/TS 120 s• THT 180s (if only THT)• H2S, MM, EM 300s• H2S, mercaptans, THT 720 s•
H2S, mercaptans, THT 900 s with CALIB for validation
of each analysis
Supervisor:
• Embedd industrial computer Windows ® based with LCD display • 128 GB of hardware storage on SSD memory
Communication: • MODBUS communication protocol• 4-20mA• Ethernet
• 3G module (optional)
Gas supply for GC operation:
• Carrier: zero air or N2 (3 bar): 4 ml/min. Use N2 if THT is present• CALIB: air or N2 50 ml/min• Sample inlet 1 bar: 80 ml/min• Pneumatic valve: 90 ml/commutation (0 ml in option)
Power supply:
• Main: 230V / 115V or 50/60 Hz • 24 V DC in option
Electrical consumption:
• 150 VA without options
Dimensions and Weight:
• Height: 1900 mm• Width: 800 mm • Depth: 600 mm • Net weight of analyzer without frame: 105 kg (175 kg with frame)
Model:Upon requestUpon request
Description :
The MEDOR ® Exd is an industrial gas chromatograph for the analysis
and monitoring of sulfur compounds in natural gas and gaseous fuels:
H2S, Mercaptans, Sulfides.
Two versions exist: ppm range or ppb range
Principle:
• Automatic sampling using a loop
• Loop injection by automatic valve on the column
• Isothermal gas chromatograph
• Detection of all compounds eluting from the column performed by
Chromatotec's wet cell sulfur specific detector
• Signal provided by electrochemical reaction between the wet cell
electrolyte and the sulfur compounds
Key points:
• Fully compliant with ASTM D 7493-08 : Standard Test Method for On -
line Measurement of Sulfur Compounds in Natural Gas and Gaseous
Fuels by Gas Chromatograph and Electrochemical Detection
• Internal automatic calibration system allowing automatic validation of
the data
• Continuous monitoring with automatic online sampling
• Analytical performances:
• Specific and very sensitive to sulfur compounds
• Results validation by automatic standard injection at each analysis
• Long term stability using wet cell detector installed in reservoir
• Extremely low maintenance
• Very long life time detector, up to 10 years including elctrolyte
• Low gas consumption, can be reduced in option
• More than 10 years data storage
• No calibration cylinders required thanks to internal calibration tube
• Automatic control with process device
• Intelligence system with tunable and interactive alarms levels
• Internal temperature and pressure monitoring
• Powerfull VISTACHROM Chromatotec ® software: • Remote monitoring & injection control• Full traceability with on board archiving of results and chromato - grams• QC Set up and control of threshold alarms• Data export by MODBUS / 4-20 mA / 0-10 V• Time stamp results• On site direct access to the analyzer with LCD screen and touch pad on front panel