Document
tsp_m52_trsMEDOR
Contenu du document
ONLINE GAS ANALYZER EXPERTS
WWW.CHROMATOTEC.COM
TRS MED O R ®
ppm (M51) or ppb (M52)
Online analysis and monitoring of odorous sulfur compounds
TSP M51 / M52
GAS ANALYZER GC 866
Chromatotec ® is specialized in VOC, Sulfur and permanent gases analysis at trace and ultra-trace levels (%, ppm, ppb, ppt). Please visit our website for more details.
Updated: February 2021
Model: M52022 (rack version) Model: M52022 (wall mounted version)
On-line applications:
Fence line
Fermentation process
Deodorization process
Fugitive emission
Targeted compounds:
Standard version: H2S / SO2 / MM / EM / DMS / DMDS / DES
Upon request: IPM / TBM / NPM / MES / 2BM / IBM / NBM / THT
Main markets:
WWTP / Refineries / Petrochemicals / Gas process sites
Ambiant / Industrial air monitoring
Odor impact management
TRS measurement
Standards:
ISO 19739:2004, DIN 51855/7
ASTM D 7493-08
Certifications:
GOST, ATEX, IECEx, CSA, CSA international
H 2
S
SO 2
Total
Reduced
Sulfur
DMS
DMDS
Me-SH
Et-SH
Odor
Index
WWW.CHROMATOTEC.COM
To contact us: sales@chromatotec.com
Chromatotec ® is continuously improving its products, therefore these specifications are subject to change without notice
NORTH AMERICAHouston - USA EUROPEBordeaux- FRANCE
ASIABeijing - CHINA
Description :
The TRS MEDOR ® is an autoGC-ED (MEDOR® Electrochemical wet cell
Detector) for the analysis and monitoring of sulfur compounds in air: H 2S,
Mercaptans, Sulfides.
Two versions exist according to levels of concentration expected:
• The TRS MEDOR ® ppm
• The TRS MEDOR ® ppb
Principle:
• Automatic sampling using a loop linked to sampling pump
• Automatic loop injection on metallic capillary column
• Isothermal gas chromatograph
• Detection of all compounds eluting from the column performed by
MEDOR® Detector: Electrochemical wet cell Detector which is SSD
Sulfur Specific Detector.
• Signal provided by reaction between the wet cell electrolyte and the
sulfur compounds
Key points:
• Automatic calibration/validation of the data, with embedded permeation
tube
• Continuous monitoring with automatic online sampling
• Analytical performances:
• Specific, linear and very sensitive to sulfur compounds
• Results validation by automatic standard injection
• Long term stability using detector installed in reservoir
• Extremely low maintenance
• Very long life time detector with electrolyte, up to 10 years
• Low gas consumption, can be reduced in option
• More than 10 years data storage
• No cylinders required thanks to internal calibration tube and gas
generators
• Automatic control with process device
• Intelligent system with tunable and interactive alarm levels
• Powerful VISTACHROM Chromatotec ® software: • Remote monitoring & injection control• Full traceability with on board archiving of results and chromato - grams• QC Set up and control of threshold alarms• Data export by MODBUS / 4-20 mA / 0-10 V• Time stamp results
TRS MED O R ®
ppm (M51) or ppb (M52)
Online analysis and monitoring of odorous sulfur
compounds
Options:
• Multiple stream selector (up to 16 streams with one analyzer)
• Modeling software with weather station for live odor/chemical impact
• Explosion proof version Exp or Exd for ATEX, IECEx zone 1 and 2
and also for CSA C1D2
• Calculation modules (Average / Statistics / Odor unit OU e/m3 and
more)
• Automatic data transfer through: Module for 4 outputs 4-20mA /
Modbus RTU or TCP IP / MGS1
• Nitrogen or air generator
• 24 V DC power supply
• Electric selection valve to reduce air consumption
• Wall mounted box XXX934
Technical Specifications:
Reduced sulfur compounds analysis:
• Speciation and Total Reduced Sulfur by sum (TRS): H2S,
SO2, DMS, MM, DES, DMDS, EM compounds
Detection limits* for H2S, SO2, DMS, MM, EM:
TRS MEDOR ® ppm LDL ? 0.1 ppm
TRS MEDOR ® ppb ? 1 ppb *more sensitive analyzers available upon request with special application
Range (adjustable depending on application):
• 0/10 or 0/100 or 0/1000 (ppb or ppm)
• Calculation: Odor Index, Total Sulfur, Total mercaptans
Relative Standard Deviation:
• RSD < 3 % on concentration over 48H
• RSD < 0.5% on retention time over 48H
Cycle Time for the following different analyses:
• H2S, SO2, MM, EM, DMS, DMDS, DES 600s• With special application following compounds can be added: IPM, TBM, NPM, 2BM, IBM, NBM, MES, THT and THT \
1200s
Supervisor:
• Embedded computer Windows ® based with LCD display
• 128 GB of Hardware storage on SSD memory
Linearity:
• > 0.995 for all compounds
Communication:
• Ethernet, remote control
Gas supply:
• Carrier: Dry air or N2 (3 bars): ? 4 ml / min.• CALIB: in continuous ~ 50 ml / min.• CALIB during validation ~ 250ml/min for M52 configuration
• Pneumatic valve 90ml/commutation
Power supply:
• Main: 230V 60Hz or 115V / 50Hz• In case of power loss, the instrument will restart automa - tically
Electrical consumption:
• 150 VA
Dimensions and weights:
To order:
TRS MEDOR ® ppb inbuilt computer 5U with CALIB
TRS MEDOR ® ppm inbuilt computer 5U with CALIB
Model:
M52022
M51022
Rack: 19’’ (5U)• Height: 222 mm • Width: 482mm• Depth: 660 mm• Net weight: 22 Kg
Wall mounted box: • Height: 800 mm • Width: 600mm • Depth: 300 mm • Net weight: 40 Kg
Entreprise(s) concernée(s) :
Produit(s) concerné(s) :
Date d'upload du document :
lundi 22 mars 2021