Document
tsp_c42_chromEnergy
Contenu du document
ONLINE GAS ANALYZER EXPERTS
WWW.CHROMATOTEC.COM
TSP C42
Model: C42022 (5U)
GAS ANALYZER GC 866
chr o mEnergy
Continuous natural gas measurement
Automated system for measurements in the range: ppm or %
Updated: December 2017
Main applications :
Heating value measurement
Natural gas composition
Propane composition
Product quality / process control
Targetted compounds and results :
C6+ / N 2 / CO 2 / C2 / C3 / IC4 / NC4 / NeoC5 / IC5 / NC5
ICV / SCV / WOBBE / Density / Relative density / Z mix / Mass molar
Main markets :
Petrochemical
Gas transport
Process
Standards:
ISO 6974-4, ISO 6976
Chromatotec ® is specialized in VOC, Sulfur and permanent gases analysis down to ultra trace levels (%, ppm, ppb, ppt). Please visit our website for more details
CO 2
CH 4
C1 to C6+
N 2
WWW.CHROMATOTEC.COM
Technical specifications:
Detection limit:
• Down to 10 ppm on Hexane
Detection range:
• ppm to %
Relative Standard Deviation: Repeatability:
• RSD < 3 % over 48 h (concentration)
• RSD < 0.5 % over 48 h (retention time)
Supervisor:
• Embeded computer Windows ® based with LCD display • 32 GB of Hardware storage on SSD memory
Communication:
• Communication protocol MODBUS• 4-20mA• Ethernet
Cycle time:
• 12 minutes for C6+ / N 2 / CO 2 / C1 / C2 / C3 / IC4 / NC4 / NeoC5 / IC5 / NC5
• 6 minutes in option
Gas supply:
• Carrier gas : He (4 bar; 1/16’’) : 18 ml/min • Calibration inlet: 1 bar; 1/16’’ • Sample inlet: 1 bar; 1/16’’: 20ml/min • Pneumatic valve 90 ml/commutation (0 ml in option)
Power supply:
• Main (230 V / 115 V 50 Hz/60Hz)
Electric consumption:
• Average: 100 VA; Peak: 200 VA
Dimensions and weight:
• Rack: 19’’ (5U)• Height: 222 mm • Width: 482 mm • Depth: 660 mm• Net weight of analyser: 22 kg
To order:
chromEnergy (integrated computer)
Cabinet for ATEX Zone 2
Cabinet for CSA Class 1 Division 2
CSA International certification
Model:
C42022
XXX060-Z-ATEX
XXX060-Z-CSA
XXX061
Description:
The chromEnergy is an industrial gas chromatograph for analysis and moni\
to -
ring of C1C6+ in natural gas and gaseous fuels by TCD detection.
Principle:
• Automatic sampling using a loop
• Loop injection by automatic valve on the column
• Back flush column for C6+ analysis and column cleaning
• Isothermal gas chromatograph
• Detection of all compounds eluting from the column performed by the
micro TCD detector
Key points:
• Continuous monitoring with automatic online sampling
• A pressure safety is located at detector outlet to protect the filament
• Calorific value unit is configurable on request
• Low maintenance instrument
• Full traceability of results and calibration, more than 10 years data
storage
• Automatic control with process device
• Intelligence system with tunable and interactive alarms levels
• Powerfull VISTACHROM ® Chromatotec ® software included in the flow
computer: • Remote monitoring & injection control• Full traceability with on board archiving of results and chromatograms• QC Set up and control of threshold alarms• Data export by MODBUS / 4-20 mA / 0-10 V• Time stamp results• Algorithms using analog and digital signals received from flowmeters, temperature, pressure and density transmitters
NORTH AMERICACHROMATOTEC Inc.18333 Egret Bay Blvd, Suite 270,Houston TX 77058 - USAPhone: +1 (281) 335 4944Fax: +1 (281) 335 4943
EUROPEAIRMOTEC AG SAS15 rue d’Artiguelongue33240 Val de Virvée, FRANCEPhone: +33 (0) 557 940 626Fax: +33 (0) 557 940 620
ASIA CHROMATOTEC Trading (Beijing) Co., Ltd.Room 1806, Building 1,Wanda Plaza, No.93, Jianguo Avenue,Chaoyang District,Beijing 100022 - CHINAPhone : +86 (0) 105 960 3283
Available options :
• Explosion proof Exp box
• ATEX Zone 2 : Ex II 3G Ex pz IIC T4
• CSA International Class 1 Div 2, group B, C & D
• Multiple stream selector (up to 16 streams with one analyzer)
• Calculation modules (average / statistics / heating value...)
• Module 4-20 mA / 0-10 V 4 output and 0 mA for instrument default
• Electric selection valve to reduce air consumption
• Gas cleaner
• 24 V DC power supply
• Carrier gas purificator
chr o mEnergy
Continuous natural gas measurement
Automated system for measurements in the range: ppm or %
To contact us: info@chromatotec.com
Chromatotec ® is continuously improving its products, therefore these specifications are subject to change without notice
Entreprise(s) concernée(s) :
Produit(s) concerné(s) :
Date d'upload du document :