TSP A25
ONLINE GAS ANALYZER EXPERTS
GAS ANALYZER GC 866
airmoVOC WMS
Monitoring of VOC in water, including BTEX, based on EPA 502.2 Method
Fields of application
- Water markets:
- Finished drinking water
- Raw source water
- Surface water
- Wastewater (head space / ppt)
- Rain water
- Process
- Finished water
- Waste water / effluent
- Air markets (option):
- Ambient air control
- Urban/Non urban area pollution control
- Indoor measurements
- BTEX/PAMS/CE analysis
Chromatotec® specializes in VOC, sulfur, and permanent gases analysis at trace and ultra-trace levels (ppm, ppb, ppt). For more details, please visit our website.
Updated: March 2021
WWW.CHROMATOTEC.COM
Principle & Features
The airmoVOC WMS uses a valve with a sample trap and a metallic capillary column. Its main features include: miniaturization, sensitivity, mobility, and flexibility. Everything from the sample port to the data storage is integrated into a wall-mounted box. It offers uninterrupted sampling with pre-concentration on an absorbent tube and a gas chromatograph with a metallic column with a programmable temperature gradient oven. Pressure control of the carrier gas is managed by a piezo-valve and each unit is tested for one week after production for quality control.
Software
Vistachrom software enables visualization and storage of data on a PC, including tools for recalculating, calibrating, exporting data, and setting up measurements. The software allows calculation of retention time, area, mass, or concentration profiles.
Purge Method
- Based on EPA 502.2 Method
- 5 ml of water sample as standard or optional 25 ml sparger
- Purge with ultra pure N2 at 40 ml/min
- Sampling time: 11 minutes
- Easy inlet calibration with a low sample volume requirement of 5 ml
- Dead volume < 15 mL
- Bubble diameter < 3 mm
- Automatic rinse
Technical Specifications
- Detection Limit: < 0.001 µg/l for BTEX
- Detection Range: 0.05 to 20 µg/l for surface water and finished drinking water
- Relative Standard Deviation (RSD): < 0.3% over 48h (Retention Time), < 3% over 48h (Concentration), < 10% for water analysis (Concentration)
- Base Line Zero Drift: < ±3%
- Linearity: R2 > 0.99 on all compounds
- Supervisor: Full result storage, Windows® based embedded computer with 10’’ TFT Color LCD display
- Gas Supply: H2 for FID and carrier gas: 30 ml/min, Air (FID): 180 ml/min, N2 (Purge): 40 ml/min
- Operation Temperature: 10 to 25°C
Options
- DET QMS for online GCMS
- airmoVOC BTEX expert 1 ppt LDL
- Automatic validation and calibration with internal CALIB
- 24 V DC power supply
- Integrated hydrogen and zero air generators
- Multiplexer: 2 to 32 streams
- 1 stream for water and 1 stream for air
- Internal or external multipoint calibration and zero with CALIB MFC, XXXCYL MFC, airmoCAL PAH
- airmoVOC C6C16 for more VOCs and S VOCs
- Analog output 4-20 mA or 0-10 V
Chromatotec® continually improves its products, hence specifications are subject to change without notice. To contact us: sales@chromatotec.com
NORTH AMERICA: Houston - USA
EUROPE: Bordeaux - FRANCE
ASIA: Beijing - CHINA
Model: airmoVOC WMS, A25022