Document
tsp_a21_airmoVOC_C6C20+
Contenu du document
ONLINE GAS ANALYZER EXPERTS
WWW.CHROMATOTEC.COM
airmoVOC C6-C20+
BTEX included
Analysis of up to 67 compounds from
C6 to more than C20: VOCs, SVOCs, halogen compounds (TO14), PAMS and PAHs.
TSP A22
Chromatotec ® is specialised in VOC, Sulfur and permanent gases analysis at trace and ultra trace levels (ppm, ppb, ppt).Please visit our website for more details.
GAS ANALYZER GC 866
Updated: July 2020
19962006
2010
Model: A22022
Tri-methylbenzene
FLUORENE
BTEX
Styrene
Naphthalene
CCL4
Gaseous PAHs
Benzo(a)pyrene
Selected by US EPA
Environment:
Urban/Non urban area pollution control Outdoor/Indoor measurementsBTEX / PAMS / CE analysisPlant / process emissions Tropical forest
Process:
Industrial Hygiene
Fenceline monitoring
Other applications:
Wastewater plant, Headspace or Purge and trap (method 502-2 or 524), Ambiant air control (PAMS and TO14)
2015
EN 14662-3
EN 15267-1
EN 15267-2
Certified
WWW.CHROMATOTEC.COM
To contact us: sales@chromatotec.com
airmoVOC C6-C20+
BTEX included
Analysis of up to 67 compounds from
C6 to more than C20:VOCs, SVOCs, halogen compounds (TO14), PAMS and PAHs.
Chromatotec ® is continuously improving its products, therefore these specifications are subject to change without notice
NORTH AMERICA
Houston - USA EUROPE
Bordeaux - FRANCE ASIA
Beijing - CHINA
Product technical specifications:
C6 to C20+ analysis:
• C6 (Hexane, Dimethylbutane) to C20+
(Benzo(a)pyrene and N-Pentacosane)
Detection limit:
• Benzene: 5 ppt or 0.0164 µg/m 3
• Naphthalene: 10 ppt or 0.05 µg/m 3
Detection range: • Benzene: 0.1 to 380 µg/m 3
• Napththalene: 0.05 to 400 µg/m 3
Relative Standard Deviation:
• < 0.3 % over 48h (RT)
• < 3 % over 48 h (Conc.)
Base Line: Zero drift:
• < 2%
Linearity:
• R 2 > 0.99 on all compounds
Supervisor:
• Full result storage (data and chromatogram)
• Embedded computer Windows® based with LCD display
• 128 GB of Hardware storage on SSD memory
• 4 USB Connecting Port
• Two RS-232 ports
• Display: 10’’ TFT Color LCD
Results:
• Data storage
Cycle time:
• 30 min to 120 min (depending on application)
Gas supply:
• H2 (FID and carrier gas): 30 mL/min (inlet 4 bars; 1/16’’
double ferrule)
• Air (FID): 180 mL/min (inlet 4 bars ; 1/8’’ double ferrule)
• Sample inlet (vacuum pump) 1/4’’ double ferrule
Operation Temperature:
• Room with air conditioning: 20 to 30°C
Sample volume:
• Up to 2200 mL (programmable)
Power supply:
• Main: 230 V / 115 V or 50 Hz / 60 Hz
• Battery: 24 V (option)
Electrical consumption:
• Mean: 300 VA, Peak 460 VA
Dimensions and weight:
• Rack: 19’’ - (5U)
• Height: 222 mm
• Width: 482 mm
• Depth: 600 mm
• Net Weight: 22 kg
Description:
The airmoVOC C6-C20+ is based on airmoVOC Mcerts certified and USEP A
selected. It is a gas chromatograph for analysis and monitoring of trace\
and
ultra trace amounts of VOCs from PAMS, TO14, TO15, and SVOCs like linear
alcanes up to N-Pentacosane or gaseous PAHs from Naphthalene to Benzo(a)
pyrene in standard with FID detector.
Miniaturization, sensitivity, mobility and flexibility are its main features.
Everything from the sample port up to the data storage is integrated in \
a 19’’-
rack 5U:
- Uninterrupted sampling with pre-concentration on 1 absorbent tube
- Gas chromatograph with 0.2 mm ID metallic capillary column and program\
-
mable temperature gradient oven and pressure / flowcontrol of the carrier gas
by piezo-valve
Principle:
• MFC for sampling controlled by Vistachrom to adapt sampling flow from
the software
• Heated injection valve
• Automatic sampling with inbuilt heated line and concentration using an
absorbent trap
• Thermo desorption from the trap and direct injection into a metallic cap\
il -
lary column regulated by a temperature gradient up to 350°C. The H2
pressure (carrier gas) at the column head is controlled by a piezo val\
ve.
• The detection of all compounds eluting from the column is performed by
a FID detector.
Key points:
• Heated sampling line up to heated valve in the GC
• The following gaseous PAHs can be analyzed: Naphthalene (C10), Me-
Naphtalene, 1-Me-Naphtalene, Acenaphtene, Acenaphtylene, Flourene,
Phenanthrene, Anthracene, Fluoranthene, Pyrene, Benzo(a)anthracene,
Chrysene, Benzo(b)fluorath, Benzo(k)fluorath and Benzo(a)pyrene (C20)
• Linear alcanes from C6 DimethylButane to C25 N-Pentacosane
• No condensation part for C6 to C20+
VISTACHROM ® software:
The software developed by Chromatotec ® allows :
• Remote monitoring
• Full traceability through archiving of results and QC
• Set up and control of threshold alarms
• Export of data MODBUS / 4-20mA / 0-10V
The software allows the calculation of retention time, area, mass or con\
cen-
tration profiles.
Options:
• Det QMS for online GC MS monitoring
• MODBUS / JBUS or MGS1 communication protocol
• Analog ouput 4-20mA or 0-10V or alarms
• Automatic validation and calibration with internal CALIB
• 24 V power supply
• Hydrogen and zero air generator for transportable analysers
• Multiplexer: up to 32 streams
• Purge module to extract VOC from water for online VOC in water analy -
sis
• Internal or external multipoint calibration and zero with CALIB MFC, XXX-
CYL MFC, airmoCAL 200 MFC
To order:airmoVOC C6-C20+ 5U inbuilt computer
Model:A22022
Entreprise(s) concernée(s) :
Produit(s) concerné(s) :
Date d'upload du document :
vendredi 19 mars 2021