brochure technique du gas analyzer gc 866
détails techniques sur l'analyseur de gaz gc 866

Contenu du document
TSP A22: ONLINE GAS ANALYZER EXPERTS
GAS ANALYZER GC 866
Benzo(a)pyrene airmoVOC C6-C20+
BTEX included
Analysis of up to 67 compounds from C6 to more than C20: VOCs, SVOCs, halogen compounds (TO14), Naphthalene, PAMS, and PAHs.
Styrene
BTEX Model: A22022
Certified EN 14662-3, EN 15267-1, EN 15267-2 2015
Environment: Urban/Non-urban area pollution control, Outdoor/Indoor measurements, BTEX / PAMS / CE analysis
CCL4 Plant / process emissions, Tropical forest
Process: Industrial Hygiene, Fenceline monitoring
Other applications: Wastewater plant, Headspace or Purge and trap (method 502-2 or 524), Ambient air control (PAMS and TO14)
FLUORENE
Gaseous Selected by US EPA 1996, 2006, 2010
Chromatotec® specializes in VOC, Sulfur, and permanent gases analysis at trace and ultra-trace levels (ppm, ppb, ppt).
Please visit our website for more details. Updated: July 2020
WWW.CHROMATOTEC.COM
airmoVOC C6-C20+
BTEX included
Analysis of up to 67 compounds from C6 to more than C20: VOCs, SVOCs, halogen compounds (TO14), PAMS, and PAHs.
Description:
The airmoVOC C6-C20+ is based on airmoVOC Mcerts certified and USEPA selected. It is a gas chromatograph for analysis and monitoring of trace and ultra-trace amounts of VOCs from PAMS, TO14, TO15, and SVOCs like linear alkanes up to N-Pentacosane or gaseous PAHs from Naphthalene to Benzo(a)pyrene in standard with FID detector. Miniaturization, sensitivity, mobility, and flexibility are its main features. Everything from the sample port up to the data storage is integrated into a 19’’-rack 5U:
- Uninterrupted sampling with pre-concentration on 1 absorbent tube
- Gas chromatograph with 0.2 mm ID metallic capillary column and programmable temperature gradient oven and pressure / flow control of the carrier gas by piezo-valve
Principle:
- MFC for sampling controlled by Vistachrom to adapt sampling flow from the software
- Heated injection valve
- Automatic sampling with inbuilt heated line and concentration using an absorbent trap
- Thermo desorption from the trap and direct injection into a metallic capillary column regulated by a temperature gradient up to 350°C. The H2 pressure (carrier gas) at the column head is controlled by a piezo valve.
- The detection of all compounds eluting from the column is performed by a FID detector.
Key points:
- Heated sampling line up to heated valve in the GC
- The following gaseous PAHs can be analyzed: Naphthalene (C10), Me-Naphtalene, 1-Me-Naphtalene, Acenaphtene, Acenaphtylene, Flourene, Phenanthrene, Anthracene, Fluoranthene, Pyrene, Benzo(a)anthracene, Chrysene, Benzo(b)fluorantt, Benzo(k)fluorantt, and Benzo(a)pyrene (C20)
- Linear alkanes from C6 DimethylButane to C25 N-Pentacosane
- No condensation part for C6 to C20+
VISTACHROM® software:
The software developed by Chromatotec® allows:
- Remote monitoring
- Full traceability through archiving of results and QC
- Set up and control of threshold alarms
- Export of data MODBUS / 4-20mA / 0-10V
Options:
- Det QMS for online GC MS monitoring
- MODBUS / JBUS or MGS1 communication protocol
- Analog output 4-20mA or 0-10V or alarms
- Automatic validation and calibration with internal CALIB
- 24 V power supply
- Hydrogen and zero air generator for transportable analyzers
- Multiplexer: up to 32 streams
- Purge module to extract VOC from water for online VOC in water analysis
- Internal or external multipoint calibration and zero with CALIB MFC, XXX-CYL MFC, airmoCAL 200 MFC
Chromatotec® is continuously improving its products, therefore these specifications are subject to change without notice. To contact us: sales@chromatotec.com