Document
tsp_a21_airmoVOC_C6C12
Contenu du document
ONLINE GAS ANALYZER EXPERTS
WWW.CHROMATOTEC.COM
airmoV O C C6-C12
BTEX included
Analysis of up to 53 compounds from
C6 to C12 and halogen compounds (TO14) or (PAMS).
TSP A21
Chromatotec ® is specialised in VOC, Sulfur and permanent gases analysis at trace and ultra trace levels (ppm, ppb, ppt). Please visit our website for more details.
GAS ANALYZER GC 866
Updated: May 2019
1996 2006 2010 Selected by US EPA
Model: A21022
Tri-methylbenzene
MTBE
BTEX
Styrene
Naphtalene
CHCL3
CCL4
Environment:
Urban/Non urban area pollution control
Indoor measurements
BTEX / PAMS / CE analysis
Plant / process emissions
Process:
Industrial Hygiene
Fence line monitoring
Other applications:
Wastewater plant, Purge and trap (method 502-2
or 524), Ambiant air control (PAMS and TO14)
2015
EN 14662-3
EN 15267-1
EN 15267-2
Certified
WWW.CHROMATOTEC.COM
To contact us: sales@chromatotec.com
airmoV O C C6-C12
BTEX included
Analysis of up to 53 compounds from
C6 to C12 and halogen compounds (TO14) or (PAMS).
NORTH AMERICAHouston - USA EUROPEBordeaux - FRANCE
ASIA Beijing - CHINA
Chromatotec ® is continuously improving its products, therefore these specifications are subject to change without notice
Product technical specifications:
C6 to C12 analysis:
• C6 (Hexane, Dimethylbutane) to C10
(Diethylbenzene, Naphtalene) and dodecane
Detection limit:
• Tri-methylbenzene: 10 ppt or 0.05 µg/m 3
• Benzene: 10 ppt in standard
1 ppt with airmoVOC Expert
Detection range: • 0.05 to 45 µg/m 3 for Tri-methylbenzene and Benzene - (A3) • 0,50 to 400 µg/m 3 for Tri-methylbenzene and Benzene - (A2) Note: Be careful with total mass and mass by compounds.
Amplification 1: On demand to decrease the sampling volume.
Relative Standard Deviation:
• < 0.3 % over 48h (RT)
• < 3 % over 48 h (Conc.)
Results:
• Data storage
• MODBUS / JBUS communication protocol (option)
Cycle time:
• 15 min, 30 min, 40 min, 60 min
Gas supply:
• H2 (FID and carrier gas): 30 ml/min (inlet 2 bars; 1/16’’ swagelock)• Air (FID): 180 ml/min (inlet 3 bars ; 1/8’’ swagelock)
• Sample inlet (vacuum pump) 1/4’’ swagelock
Sample volume:
• 30 to 700 ml (programmable)
Power supply:
• Main: 230 V / 115 V or 50 Hz / 60 Hz
• Battery: 24 V (option)
Electrical consumption:
• Mean: 150 VA, Peak 360 VA
Dimensions and weight:
• Rack: 19’’ - (5U) • Height: 222 mm• Width: 482 mm• Depth: 600 mm• Net Weight: 22 kg
Principle:
The airmoVOC C6-C12 uses a valve with 1 sample trap, and a metallic
capillary column .
• Miniaturization, sensitivity, mobility and flexibility are its main features.
Everything from the sample port up to the data storage is integrated in
a 19 ’’-rack 5U :
• Uninterrupted sampling with pre-concentration on 1 absorbent tube
• Gas chromatograph with 0.2 mm ID metallic column and program -
mable temperature gradient oven and pressure / flowcontrol of the
carrier gas by piezo-valve
Before delivery the analyser is tested for one week by our quality contr\
ol
department.The Vistachrom software enables the user to visualize and store
data on a PC. Furthermore it enables the user to recalculate, calibrate \
the
instrument export data and to set-up measurement.
The software allows the calculation of retention time, area, mass or
concentration profiles.
Options:
• On-line results are transmitted via: MODBUS / JBUS or MGS1
communication protocol
• AirmoVOC Expert 1 ppt
• Automatic validation and calibration with internal CALIB
• 24 V power supply
• Hydrogen and zero air generator for transportable analysers
• Multiplexer: 2 to 32
• Purge module to extract VOC from water for online VOC in water
analysis
• Internal or external multipoint calibration and zero with CALIB MFC,
XXXCYL MFC, airmoCAL MFC or airmoCAL D
• Up to C16 with airmoVOC C6C16
To order:airmoVOC C6-C12 5U inbuilt computer
Model:A21022
For more information, consult our Case Study
Ambient air monitoring like :
airmOzone 88, airmOzone + DET QMS 96 from PAMS-TO14,
Car painting line, airmoTWA, airmoVOC BTEX
Entreprise(s) concernée(s) :
Produit(s) concerné(s) :
Date d'upload du document :
vendredi 19 mars 2021