MCA 100 BIO
MULTICOMPONENT ANALYZER FOR BIOGAS
The MCA 100 Bio series of gas analyzers by ETG it’s the ideal solution for biogas measurement and
industrial combustion applications because of their accuracy, stability, reliability, broad
measurement range, and the variety of available form factors.
Unlike other analyzers, ETG MCA 100 non ?dispersive infrared (NDIR) gas analyzers measure multiple
gases in an instrument with a single optical path platform.
Single ?gas analyzers are inadequate when using methane as a biofuel because the gas often contains
large amounts of CO as a contaminant. ETG analyzers have the ability to measure CO2, CO, and O2 in
addition to methane and therefore
provide the optimal combination of gases for combustion
optimization.
The enhanced optics and electronics of our NDIR analyzers have virtually eliminated zero drift after
the initial warm up period. The temperature and pressure compensation eliminates the major causes
of span drift in many NDIR instruments.
• Analysis of CO 2, CH 4, O 2, H 2S
• NDIR technology for CO
2 and CH 4
•
Automatic air purge for long life H 2S
cell
• Arm Processor
• Touch Screen monitor
• Ethernet and USB Remonting •
Analog & Digital Signal Output
• Low cost ownership
• 19 “ Rack mounting or Wall mounting
version
• Modbus, Profibus, Ethernet ( optional )
• Autozero
• Data downloadable on SD Card
T90 & T 10<20
seconds
+/?5 ppm
abs or +/ ?
3.0% rel
+/?5 ppm
abs or +/ ?
3.0% rel 0?2000 ppm
0 ?5000 ppm 1 ppm Hydrogen
Sulfide Electro?
chemical
sensor
< 40 seconds
from
ambient to
0.15 O2 +/?0,1% abs
or +/ ? 1.5%
rel
+/?0,1% abs
or +/ ? 3.0%
rel
1.01 ?25.00% 0,01% Oxygen Electro?
chemical
sensor
T90 & T 10<10
seconds
+/?0,7%
+/ ? 0,1% abs
or +/ ? 0.8%
rel
+/ ?0,8 F.S.
+/ ? 0,3% abs
or +/ ? 3.0%
rel
+/ ? 5% rel 0?50%
0.00 ?16.00%
16.01?20.00%
0,01% Carbon
Dioxide NDIR ( Non ?
Dispersive
Infrared )
T90 & T 10<10
seconds +/?0,8% +/?1% F.S.
+/ ? 2% rel
+/ ? 1% rel
+/ ? 5% rel 0?100%
201ppm ?40%
40 ?80%
80 ?100% 0,01% Methane NDIR ( Non ?
Dispersive
Infrared )
Time
Precision
Accuracy
Range
Resoluti
on
Gas
Measurement
Method
4
?20 ma for each measured compound Analog Signal Output
Ethernet – Profibus –Modbus Communications (optional)
?300 to 3.000 m ( ? 1.000 to 10.000 ft ) Operating Altitude
To 95% RH ( Non ?condensing)
°F) Operating Temperature
30 seconds ready,3 minutes useable,30 minutes to full performance Warm?up Time
1 second Data Refresh Rate
Response times are specified at a sample flow rate of a 1 liter per minute
through the MCA 100 sample cell Response Time
MCA 100 BIO MULTICOMPONENT ANALYZER FOR BIOGAS
SPECIFICATIONS
Analog Out 6
Digital Out 5
Vent 4
Exaust 3
Purge In 2
1
MECHANICAL DIMENSIONS
PNEUMATIC/ELECTRIC
1 2 3 4
5 6
Other Version of Biogas Analyzers
• Wall mounting Biogas analyzer version model 6500
• Portable Biogas analyzer model 3200 P ETG RISORSE E TECNOLOGIA
Via P. Carpignano 23
14026 Montiglio M.to (AT) ITALY
Tel. +39 0141 99.4905 Fax 99.49.71
sales@etgrisorse.com
www.analisi-biogas.com
www.etgrisorse.com
15.75 kG Weight
132,5 x 495,0 x 430,5 mm Mechanical dimensions
Rapid fittings 6.0 OD 4.0 ID Pneumatic connection
from 100 to 240 Vac 47 ?63 Hz Electrical supply
Zero & span user selectable. Automatic calibration ( optional)
Monitor
Configurable for gas concentration, analyzer fault, high temperature , etc.. Digital Output