Document
fiche technique du cleaning module-ii pour systèmes de surveillance ami
fiche technique pour le module de nettoyage chimique automatique

Contenu du document
Swan Analytische Instrumente AG
CH-8340 Hinwil, Switzerland
Tel. +41 44 943 63 00
Email: swan@swan.ch
Website: www.swan.ch
Cleaning Module-II
Data sheet no. DenA82312000
Optional module for the automatic, chemical cleaning of flow cell and photometer. To use conjointly with the AMI Codes-II, AMI Codes-II CC, AMI Codes-II TC, AMI Phosphate-II, AMI Phosphate HL or AMI SAC254 monitoring systems.
Ensuring highest accuracy of measurement:
- Chemical cleaning is achieved with periodical additions of one or two cleaning solutions, e.g., 2.5% hypochlorite solution (Javel water) and/or diluted sulfuric acid (2.5%).
- Counteracts bio-growth inside the flow cell and photometer and thereby ensures reliable measured values.
- Factory tested and ready for installation.
User-friendly operation:
- Automatic level monitoring of the reagents.
- Cleaning interval is controlled by the AMI transmitter and can be programmed individually.
- During the cleaning process the AMI transmitter stays in hold mode.
Scope of delivery:
- Complete system installed on mounting panel including control electronics, reagent dosing system with switching valves, reagent containers and a sensor for level monitoring of the reagents.
Technical data:
- Panel made of PVC Dimensions: 170 x 850 x 200 mm
- Power supply via AMI transmitter.
Order information
Order no.: Cleaning Module-II A – 8 2 . 3 1 2 . 0 0 0
Accessories:
- A-86.191.042 Tubing set for Cleaning Module-II
- A-86.191.151 Frame with tube 2.29 mm for the PeriClip pump
06/2020 Subject to change without notice