Document
brochure-enviroFlu HAP-TriOS-BIONEF
Contenu du document
1
enviroFlu
30SXXXXX0
PAH, oil-in-water using UV fluorescence
enviroFlu-HC is the new generation of immersion sensors
for measurement of oil-in-water. The used measuring
principle of UV fluorescence is much more sensitive than the
conventionally used infrared scattering or absorption method.
This allows to determine even the slightest traces of PAH's, for
example in drinking water and cooling water condensates.
Application areas include the petrochemical industry,
leakage detection in cooling and wastewater streams as
well as environmental monitoring. The devices enable both
stationary use in shafts, flows or piping, and mobile use
through an optional hand-held measuring instrument. An
innovative coating reduces fouling of the optical measuring
window and minimizes the maintenance.
Benefits
• Without sampling and preparation of test samples
• Real time sensor
• Without reagents
• High sensitivity and selectivity
• Optical window with nano coating
Applications
• Drinking water
• Wastewater
• Airports
• Cooling water
• Desalination plants
• Refineries
• Pipeline monitoring
• Bilge water monitoring
• Exhaust gas cleaning with approval for ship use
according to IMO regulation MEPC.184(59)
Naphtalene Acenaphtylene Acenaphtene Fluorene Phenanthrene Anthracene Fluoranthene Pyrene Benzo[a]anthracene
Crysene Benzo[b]fluoranthene Benzo(k)fluoranthene Benzo[a]pyrene Dibenzo(a,h)anthracene Ideno(1,2,3-c,d)pyrene Benzo(g,h,i)perylene
enviroFlu
D02-032en201903 brochure enviroFlu
TriOS Mess- und Datentechnik GmbH ? B?rgermeister-Br?tje-Str. 25 ? D-26180 Rastede ? Germany fon: +49 (0) 4402 69670 - 0 ? fax: +49 (0) 4402 69670 - 20 ? in\
fo@trios.de ? www.trios.de
2
Technical Specifications
Measurement
technology
light source Xenon flash lamp + filter (254 nm)
detector Photo diode + filter (360 nm)
Measurement principle Fluorescence
Parameter PAH, oil
Measuring
range
enviroFlu-HC
500
PAH: 0...50 ppb, 0...500 ppb
Oil: 0...1.5 ppm, 0...15 ppm typical
enviroFlu-HC
5000
PAH: 0...500 ppb, 0...5000 ppb
Oil: 0...15 ppm, 0...150 ppm typical
Detection limit enviroFlu-HC 500 0.3 ppb
enviroFlu-HC 5000 0.5 ppb
Measurement accuracy ± 5 % FS
Reproducibility ? 0,5 % FS
Turbidity compensation No
Data logger No
T100 response time ? 10 s
Measurement interval ? 5 s
Material Housing Stainless steel (1.4571/1.4404) or titanium (3.7035)
Head POM black with synthetic quartz glass
Dimensions (L x Ø) 311 mm x 68 mm ~12.2? x 2.6?
Weight stainless steel ~ 2.7 kg ~ 6 lbs
titanium ~ 1.9 kg ~ 4.2 lbs
Interface digital RS-232 (TriOS)
analog 4...20 mA, 0...5 V
Power consumption ? 3.5 W
Power supply 12...24 VDC (± 10 %)
Maintenance effort ? 0.5 h/month (typical)
Calibration/maintenance
interval 24 months
System compatibility Analog Out (0...5 VDC, 4...20 mA)
Warranty 1 year (EU: 2 years) US: 2 years
enviroFlu
D02-032en201903 brochure enviroFlu
TriOS Mess- und Datentechnik GmbH ? B?rgermeister-Br?tje-Str. 25 ? D-26180 Rastede ? Germany fon: +49 (0) 4402 69670 - 0 ? fax: +49 (0) 4402 69670 - 20 ? in\
fo@trios.de ? www.trios.de
3
INSTALLATION
Max. pressure
with SubConn 30 bar ~ 435 psig
with fixed cable 3 bar ~ 43.5 psig
in FlowCell 1 bar, 2...4 L/min ~ 14.5 psig, 0.5 to 1 gpm
Deepsea version 600 bar
Protection type IP68 NEMA 6P
Sample temperature +2...+40 °C ~ +36 °F to +104 °F
Ambient temperature -5...+55 °C
(2...+40 °C for specified accuracy)
~ +23 °F to +131 °F
(~ 35,6 °F to 104 °F for specified accuracy)
Storage temperature -20...+80 °C ~ -4 °F to +176 °F
Inflow velocity 0.1...10 m/s ~ 0.33 fps to 33 fps
Max. immersion depth
300 m with SubConn-8pin underwater
plug
~ 984.25 ft with SubConn-8pin underwater
plug
30 m with fixed cable ~ 98.43 ft with fixed cable
optional: 6000 m Tiefsee Version optional: ~ 19685.04 ft deepsea version
enviroFlu
D02-032en201903 brochure enviroFlu
TriOS Mess- und Datentechnik GmbH ? B?rgermeister-Br?tje-Str. 25 ? D-26180 Rastede ? Germany fon: +49 (0) 4402 69670 - 0 ? fax: +49 (0) 4402 69670 - 20 ? in\
fo@trios.de ? www.trios.de