Document
App Note T005 - T920 Redox - T5 SO2 in Wine RevB
Contenu du document
Free and Total Sulfur Dioxide (SO
2
) in
wine (Ripper method) by
automatic titration
Required reagents and solutions
• Titrant: Purchased or prepared iodine/iodide titrant,
0.01M (0.02N)
• Standardizing Solution (optional): Purchased or prepared
0.01M (0.01N) standard sodium thiosulfate solution
• Reagent Solutions: Purchased or prepared 1N Sodium
Hydroxide, purchased or prepared 25% Sulfuric Acid (1 +
3 H
2SO 4), prepared conditioning solution
• Solid Reagent: Sodium Bicarbonate
• Reagent grade water (RGW)
Use suitable Personal Protective Equipment (PPE) and
ventilation as recommended by the Safety Data Sheets
(SDS) for the chemicals utilized during this procedure.
Water Analysis Instruments,
Thermo Fisher Scientific
Key words
Sulfite, sulfurous acid, SO 2, juice, must, iodine, redox, Orion
9770BNWP, Orion Star T920, Orion Star T940, titrator.
Preprogrammed methods
T5 SO 2 Wine
Introduction
Free and Total sulfur dioxide in wine are determined using
the preprogrammed method, T5 SO
2 Wine. To determine
free sulfur dioxide, the sample is treated with appropriate
reagents and titrated to the equivalence point using iodine/
iodide titrant. To determine total sulfur dioxide, the sample
is reacted with alkaline solution for a set period of time,
then treated with appropriate reagents and titrated to the
equivalence point using iodine/iodide titrant.
Recommended equipment
• Thermo Scientific ™ Orion Star ™ T920 Redox Titrator
or T940 All-In-One Titrator or equivalent with a 20 mL
burette
• Thermo Scientific
™ Orion ™ 9770BNWP platinum/iodide
(chlorine) electrode or equivalent
• Glass pipette or graduated cylinder, 10 mL graduated
• 25 or 50 mL graduated cylinder
• 100 and 150 mL beakers
No. T5
APPLICATION NOTE
Titrator setup
Connect the electrode and the stirrer probe to the
titrator. If not previously done, import the T5 SO
2 Wine
preprogrammed method into the titrator from the Methods
screen
1. Rinse and fill the burette with titrant. See the
titrator user manual for details on setting up the titrator.
If bubbles are visible in the tubing, dispense titrant (from
the Burette screen) until the bubbles have been expelled.
Tap the tubing to dislodge bubbles. Consider standardizing
the titrant before titrating samples. See Titrant section
b e l ow.
T5 SO
2 Wine method: Preprogrammed parameters
ElectrodeParameter
Electrode Type Redox
Electrode Name Edit as desired
TitrantParameter
Titrant Name I2
Titrant ID Edit as desired
Conc Input Mode Standardization
Nominal Concentration 0.01 M
2
Standardize TechEquivalence Pt.
Number of Endpoints 1
Results Units M
Standardize Reaction Ratio 2
Standard Name Na
2S2O3
Standard AmountFixed volume, 10 mL
Standard Concentration 0.010 M
3
Pre-dose Titrant Volume4 mL
Max total titrant volume 10 mL
Stand. Process Control User defined
?E 10 mV
?Vmin 0.02 mL
?Vmax 0.100 mL
dE/dt 10
t min 2 seconds
t max 8 seconds
Threshold 250
Pre-stir Duration 5 sec
Stir Speed Medium
Titration Parameter
Titration Technique Equivalence Pt.
Number of Endpoints 1
Titration Type Direct
Blank Required No
Result Units mg/L
Reaction Ratio 1
Sample Amount Fixed vol., 25 mL
Sample MW 64.07
Pre-dose Titrant Volume 0.1 mL
Max total titrant volume 10 mL
Titration (Contd.)Parameter (Contd.)
Titration Process Control User defined
?E 5 mV
?Vmin 0.020 mL
?Vmax 0.100 mL
dE/dt 10 mV/min
t min 2 seconds
t max 4 seconds
Threshold 250
Pre-stir Duration 10 sec
Stir Speed Medium
Sample ID Manual
Electrode preparation
Before the first titration of the day, place electrode into
conditioning solution for 10 or 15 minutes. Rinse thoroughly
with RGW before and between titrations. At the end of the
day, clean any foreign materials from the platinum sensor
by wiping with a moistened lint-free wiper. Thoroughly rinse
the electrode with RGW and store the electrode dry.
Reagent solution preparation
Prepared 25% Sulfuric Acid: Place 750 mL of RGW
in a large beaker or graduated cylinder. Add 250 mL of
concentrated (98%) H
2SO 4. Mix and allow to cool. Store in
a sealed container at room temperature.
Conditioning solution: Dispense 0.5 mL of titrant into a
150 mL beaker and add RGW to the 100 mL mark.
Sample preparation
Free SO 2: Accurately add 25.0 mL of wine to a beaker.
Add RGW to the 60 mL mark on the side of the beaker.
Proceed to Sample Titration.
To t a l S O
2: Accurately add 25.0 mL of wine. Add 25 mL of
1 N NaOH to a beaker. Mix and allow 10 minutes reaction
time. After 10 minutes, add RGW to the 60 mL mark on the
side of the beaker. Proceed to Sample Titration.
Sample titration
1. Free SO 2
a. From the Methods screen, select option to run the
saved method T5 SO
2 Wine.
b. Rinse the electrode, stirrer, and dispenser with RGW.
Place the electrode, stirrer, and dispenser into the
prepared sample in the beaker. Ensure that the
dispenser tip is inserted below the surface of the
sample. Ensure that the reference pellet on the side of
the electrode is immersed.
c. Add 5 mL of 25% sulfuric acid to the beaker. Start
the titration immediately. When the stirrer starts to
spin, immediately add about 1 g (¼ teaspoon) of solid
Range
This preprogrammed titration method covers a range of
about 25 to 250 mg/L of SO
2, when using 0.01 M (0.02 N)
Iodine/Iodide titrant and the prescribed volume of sample.
Method modifications
• For samples of a concentration less than 25 mg/L:
– Double the sample volume tested, e.g., 50 mL.
– Edit the method (SO
2 Wine), edit Titration, change
Sample Amount to Variable Volume, and Save. When
prompted by the titrator, enter 50 mL as the sample
volume.
• For samples of a concentration of greater than 250 mg/L,
choose one option:
– Use a smaller sample volume, e.g., 15 mL. Edit the
method (SO
2 Wine), edit Titration, change Sample
Amount to Variable Volume, and Save. When prompted
by the titrator, enter 15 mL as the sample volume.
• For shorter titrations:
– Use a pre-dose to shorten the analysis time.
– Edit the method (SO
2 Wine), edit Titration, change the
Pre-dose Titrant Volume, and Save.
– If desired, copy the method to create two methods:
one method for free SO
2 and another method for total
SO
2. Adjust the predose as desired for each method.
– In general, a pre-dose that can be expected to work
well will be a volume of about 1 mL less than the
expected final volume of the titration.
– For example, a sample with a concentration near
100 mg/L will take about 4 mL of titrant. To make the
titration go faster, edit the method to use a predose of
3 mL.
bicarbonate to the beaker. The sample will bubble and
foam. Addition of the bicarbonate may be omitted if
results are satisfactory without it.
d. Results are reported as mg/L SO
2.
2. To t a l S O
2
a. From the Methods screen, select option to run the
saved method T5 SO
2 Wine.
b. Rinse the electrode, stirrer, and dispenser with RGW.
Place the electrode, stirrer, and dispenser into the
prepared sample in the beaker. Ensure that the
dispenser tip is inserted below the surface of the
sample. Ensure that the reference pellet on the side of
the electrode is immersed.
c. Add 10 mL of 25% sulfuric acid to the beaker. Start
the titration immediately. When the stirrer starts to
spin, immediately add about 1 g (¼ teaspoon) of solid
bicarbonate to the beaker. The sample will bubble and
foam. Addition of the bicarbonate may be omitted if
results are satisfactory without it.
d. Results are reported as mg/L SO
2.
Results
Parameter SampleAverage
(n = 3) SD
(mg/L) Analysis
Time
Free SO
2 White wine 26.3 mg/L
FSO 2 0.6 02:39
minutes
Free SO
2 Red wine 29.9 mg/L
FSO 2 0.7 03:03
minutes
Total SO
2 White wine 65.9 mg/L
TSO 2 0.8 03:49
minutes 4
Total SO 2 Red wine 51.8 mg/L
TSO 2 0.9 02:51
minutes 5
Free SO 2 in white wine
Total SO
2 in white wine Total SO 2 in red wine
Free SO
2 in red wine
Titrator and electrode care
Refer to the titrator and electrode user manuals for details on cleaning, storage, and maintenance recommendations to
keep the titrator and electrode performing well. Main points for care are summarized below.
Daily Care Weekly or Biweekly Care As Needed
• If bubbles are visible in the titrator
tubing, dispense titrant until bubbles
have been expelled. Tap tubing to
dislodge bubbles that stick.
• At the start of the testing, immerse the
electrode in diluted titrant conditioning
solution for 10 or 15 minutes.
• Rinse electrode well with RGW before
and between titrations.
• Clean any foreign materials from the
platinum sensor by wiping gently with
a moistened lint-free wiper.
• Storage: Thoroughly rinse the electrode
with RGW and store the electrode dry. •
Consider standardizing the titrant on
a weekly basis, or more frequently, as
desired.
• Prepare a small batch of conditioning
solution for the week. •
For slow or drifty electrode response,
soak 15 minutes in warm 1% laboratory
detergent while stirring. Gently brush the
platinum sensor and reference pellet with
a soft toothbrush. Rinse well with RGW
afterward.
• If still slow or drifty, use Orion pH
cleaning solution C, immersing only the
platinum sensor. Do not immerse the
reference pellet. Rinse well, then soak in
conditioning solution solution 10 or 15
minutes before use.
• See the electrode user manuals for
maintenance details.
Notes
1Refer to the user manual for detailed instructions.
2If the concentration of the iodine titrant is listed in normality (N), divide by 2 to obtain the molarity (M). For example, 0.02N
iodine ÷ 2 = 0.01M iodine.
3If the standardizing solution concentration is not 0.0100 M (0.0100 N), enter the true value.
4Using a predose of 1.0 mL. For a shorter run, try a predose of 1.5 or 2.0 mL.
5Using a predose of 1.5 mL.
Titrant
Over time, standard titrant solutions age and can change concentration. For higher accuracy, determine the exact
concentration by standardizing the titrant. It is common to standardize on a weekly basis, but other standardization
frequencies may be suitable.
1. Standardizing titrant
a. Pipet 10.0 mL standardizing solution, 0.01M (0.01N) standard sodium thiosulfate, into a clean 100 mL beaker. Add
about 60 mL of RGW to the beaker.
b. Select the SO
2 Wine method. At the pre-titration screen, select the standardize option.
c. Add 2 drops of 25% sulfuric acid to the beaker (to lower the pH to < 5).
d. Start the titration.
e. The new standardized titrant concentration will automatically be saved and used for subsequent SO
2 Wine method
titrations.
2. Certified standardized titrant solutions
a. Some customers may prefer not to standardize their titrant, instead choosing to purchase and use certified
standardized titration solutions. In this case, edit the Titrant section of the method and enter the certified concentration
and titrant ID (i.e., lot number, if desired).
To purchase Thermo Scientific laboratory products, please contact your local equipment distributor and reference the part
numbers listed below:
Product Description Cat. No.
TitratorsThermo Scientific Orion Star T920 pH Titrator without electrode
START9200
Thermo Scientific Orion Star T940 All-In-One Titrator without electrode START9400
Electrodes Thermo Scientific Orion Platinum/Iodide (chlorine) Electrode 9770BNWP
Accessories 100 and 150 mL beakers
25 or 50 mL graduated cylinder
10 mL graduated pipet
Reagent Grade Water Thermo Scientific
™ Barnstead ™ Smart2Pure ™ 12 UV Water Purification System 50129890*
Reagents 0.01M (0.02N) iodine/iodide titrant
0.01M (0.01N) standard sodium thiosulfate solution
1M Sodium Hydroxide
Sulfuric Acid, 98% or 25%
Sodium Bicarbonate, solid
*Please contact your local Thermo Scientific representative for support on ordering the best water purification system for your application. And visit our
website at www.thermofisher.com/T900titratorseries.
References
Zoecklein et al. Wine Analysis and Production , Chapman and Hall. 1995.
Iland et al. Chemical Analysis of Grapes and wine: techniques and concepts . 2
nd Edition, 2013.
Find out more at thermofisher.com/T900titratorseries
Australia: (613) 9757-4300 In Australia: (1300) 735-295 China: (86) 21-6865-4588
Germany: (49) 6184-90-6321 India: (91) 22-6716-2261/2247 Japan: (81) 045-453-9175
North America: 1-978-232-6000 Toll Free: 1-800-225-1480 Singapore: (65) 6778-6876
©2019 Thermo Fisher Scientific Inc. All rights reser ved. All trademarks are the property of Thermo Fisher Scientific
and its subsidiaries unless otherwise specified. AN-SO2WINE-E 0319 RevB
Entreprise(s) concernée(s) :
Question(s) liée(s) :
Date d'upload du document :
mardi 21 juin 2022